Team:UNIPV-Pavia/Methods Materials/Evaporation
From 2009.igem.org
(→The "frame effect" on the plate) |
(→The "frame effect" on the plate) |
||
Line 34: | Line 34: | ||
We tried to correct this phenomen dispensing water through the instrument injector, but we did not find a satisfactory dispensing protocol, as you can see in the figure below, which shows the final volumes in the wells after a 24 hours incubation. The initial volume was 200 μl (LB medium) and others 110 μl were added during the incubation. | We tried to correct this phenomen dispensing water through the instrument injector, but we did not find a satisfactory dispensing protocol, as you can see in the figure below, which shows the final volumes in the wells after a 24 hours incubation. The initial volume was 200 μl (LB medium) and others 110 μl were added during the incubation. | ||
- | You can find more informations about the experiment at [http://aimed11.unipv.it/iGEM2009 Download Protocol] - Test n.3, 10/06/09. | + | You can find more informations about the experiment at [http://aimed11.unipv.it/iGEM2009/TestTECANInfiniteF200/10_06_09%20Test3/ Download Protocol] - Test n.3, 10/06/09. |
Line 48: | Line 48: | ||
- | You can find more informations about the experiment in the figure on top at [http://aimed11.unipv.it/iGEM2009 Download Protocol] - Test n.11, 21/07/09. | + | You can find more informations about the experiment in the figure on top at [http://aimed11.unipv.it/iGEM2009/TestTECANInfiniteF200/21_07_09%20Test11/ Download Protocol] - Test n.11, 21/07/09. |
Revision as of 11:25, 19 October 2009
|
Measurements - Evaporation |
|
The "frame effect" on the plateIn the first experiments we did not use the lid on the microplate and we find out that the wells were subjected at a non uniform evaporation phenomenon, if the plate was left in incubation at 37°C inside the instrument. In particular we observed that the evaporation was larger in the wells on the frame of the plate. We tried to correct this phenomen dispensing water through the instrument injector, but we did not find a satisfactory dispensing protocol, as you can see in the figure below, which shows the final volumes in the wells after a 24 hours incubation. The initial volume was 200 μl (LB medium) and others 110 μl were added during the incubation. You can find more informations about the experiment at [http://aimed11.unipv.it/iGEM2009/TestTECANInfiniteF200/10_06_09%20Test3/ Download Protocol] - Test n.3, 10/06/09.
|